20 MAINTENANCE AND
INSPECTION OF THE
SJ-410
20.1 Daily Care
After a measurement task has been completed, store the SJ-410 components in the
attache case to keep out dust and moisture. Before storing the drive/detector unit, check
the condition of the SJ-410 for smooth start of the next task.
■ Checking for normal operation
To judge whether the SJ-410 is in normal operation, after calibrating it with the supplied
roughness specimen (Order No.178-601, 178-605) check that the dispersion of Ra values
is within
±0.05 µm, which are obtained from repeated measurements of the same point. This
dispersion in the roughness specimen is a value obtained under the conditions that there
are no dent and abrasion on the detector stylus tip and no scratch and abrasion on the
specimen surface.
NOTE
When the measured point on the supplied roughness specimen (Order No.178-601,
178-605) is changed during repeated measurement, the dispersion of ±0.09 µm (3.543
µin) (±3% of the nominal value) included in the roughness specimen is added to that of
the Ra values. Care should be exercised.
■ Storing the touch pen in the display unit
Store the touch pen in the drive unit housing on the display unit.
■ Cleaning the surface of the SJ-410
If the SJ-410 is soiled, wipe it using a soft, dry cloth. Do not use thinner or benzene for
cleaning.