Example of 1 profile using evaluating condition B
Evaluation
start point
Pre-
tr
a
v
e
l
Post-
travel
Sampling lengths 1 to 7
Traverse length
Traverse length: 6.9 mm (0.271 in)
Approach travel (0.5 mm) (0.019 in) + Pre-travel length λc/2 (0.4 mm) (0.015 in) +
Evaluation length λc (0.8 mm) (0.031 in) × 7 sampling lengths + Post-travel length λc/2
(0.4 mm) (0.015 in)
Switching between the evaluating conditions A and B
Touch
or to switch between the screens.
Switching between the Evaluation Condition Menu screens
The factory-set default of evaluating condition B is set to “OFF”. To calculate using 2
different evaluating conditions, switch to Evaluating Condition Menu screen B and enable
B by touching
on the screen. To calculate using 1 evaluating condition, touch
to disable B.
.
Disabled Enabled
Evaluating Condition B Menu screen