13. OPERATING ENVIRONMENT SETUP
No. 99MBB765A
13-21
Calibration Setup screen
Touch
.
To cancel the input operation, touch
.
The entered nominal value
is displayed on the Calibration
Setup screen.
Calibration Setup screen
3 Press to begin measurement.
After measurement, the pitch result is displayed.
To cancel the displayed result, touch
[Cancel].
4
2 measurements need to be made, at 0.5 mm/s and 0.1 mm/s
(0.019 in/s and 0.003 in/s).
Waveform Display screen
Touch [Evaluation profile] to display the calibrated
waveform.
TIP
For information about enlarging/reducing or scrolling
waveform, refer to "4.6.2.Display the evaluation
profiles" (page 4-18).
Calibration Setup screen
5
Touch [Update].
Calibration Setup screen
The calibration result is changed.