15. PRELIMINARY MEASUREMENT (AUXILIARY FUNCTIONS)
No. 99MBB765A
Cross-travel Table Adjustment screen
After the completion of measurement, retract the stylus and touch
.
Cross-travel Table Adjustment screen
Turn the tilt knob of the cross-travel table in the plus direction only
by the amount of the correction value, and touch
.
NOTE
• The plus direction and minus direction of the correction
value vary depending on the tilt.
Cross-travel Table Adjustment screen
Adjustment has been completed.
Touch
[Close] to return to the Preliminary Measurement
screen.
NOTE
• Perform leveling operation with the cross-travel table again to make sure that leveling
operation has done properly. To level the measuring surface more precisely, repeat
leveling operation with the cross-travel table.
IMPORTANT
• The rotational direction of the tilt knob In the case where the tilt knob is located in the
measurement direction of the 50 mm (1.968 in) drive unit (the right side from the
anterior view), the rotational direction of the tilt knob is in this direction. However, the tilt
knob rotates in the reverse direction when the micrometer head is located in the
direction opposite the measurement direction of the 50 mm (1.968 in) drive unit (the left
side from the anterior view).