■ Attaching the tilt adjustment unit, X-axis adjustment unit, and automatic positioning unit
NOTE
• Make sure to use the supplied screws.
• When attaching the simple column stand, exercise care so as not to drop the drive unit
or unit to prevent damage.
• When the units are used in combination, the straightness accuracy of the SJ-411/412
drive unit is reduced for approx. 0.2 μm (7.874 μin).
1 Remove the mount board from the simple column stand.
Simple column stand
Mount board
Mounting screws (M4 x 8)
2 Remove the simple column stand mount board from the automatic positioning unit
and X-axis adjustment unit.
Simple column stand mount board
X-axis adjustment unit
Automatic positioning unit
Simple column stand mount board
3 Loosen the 2 securing screws to remove the mount board from the tilt adjustment
unit.
Securing screws (M4 x 8)
Tilt adjustment unit
Securing screws
(M4 x 8)
Mount board
No. 99MBB765A