3. SETTING UP THE SJ-410
No. 99MBB765A
3-11
■ Detaching the detector
NOTE
Be sure to detach the detector from the drive unit in the detector origin position. If the
connector is not in the origin position, it is not possible to detach the detector, since the
connector of the drive unit is deep inside the unit.
1 Move the detector to the origin position.
TIP
For information about moving the detector to the origin position, refer to "5.2.2
Repositioning the drive unit for narrow-area measurement" (page 5-3).
2 Turn the SJ-410 off.
3 With the drive unit at the origin point position, quietly pull out the detector from the
drive unit.
Detaching the detector
Drive unit