Jitter analysis and clock data recovery
R&S
®
RTP
1027User Manual 1337.9952.02 ─ 12
Serial
standard
Nom. bit
rate
Bandwidth Rel. bandwidth Order Damping
Thunder-
bolt 10G
10.3125
Gbps
1667 6.19 MHz 1st -
XAUI 3.75 Gbps 1667 2.25 MHz 1st -
Remote command:
ADVJitter<m>:SIGNal:SERStandard on page 2267
Signal Type
Select, if the analyzed signal is a data or clock signal.
Remote command:
ADVJitter<m>:SIGNal:TYPE on page 2262
Quick start analysis
Starts a quick advanced jitter analysis.
Before starting the quick start analysis, set the Analyzed source and the Signal Type .
The quick start analysis performs the following:
●
A bit rate estimation is performed on the applied signal.
●
A software CDR of second order is used as reference signal for the jitter analysis.
●
Enables basic measurement "Event Count" and advanced jitter components:
TJ@BER, RJ, TJ, DJ and PJ.
●
In the "Result" dialog it enables:
– "Bathtub" for total jitter @ BER (TJ@BER).
– "Step response" for the analyzed source signal.
– "Histogram" for RJ, TJ, DJ, DDJ and PJ.
●
Enables the advance jitter measurement analysis.
Remote command:
ADVJitter<m>:SIGNal:QUICkmeas on page 2262
Reference Level
Opens a dialog to set the reference levels. For details, see Chapter 8.2.4, "Reference
levels", on page 342.
Gating
Opens a dialog to set up the gate of the waveform for the advanced jitter analysis. The
gate limits the measurement to a user-defined range of the waveform.
Advanced jitter and noise (option R&S RTP- K133/K134)