Remote control commands
R&S
®
RTP
1286User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
<n> 1..4
Selects the complex pattern condition.
Parameters:
<Enable> ON | OFF
*RST: OFF
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:FRPTtern <PatternFrom>
Sets the pattern or a start value of a pattern range.
Suffix:
<m>
.
1..3
<n> 1..4
Selects the complex pattern condition.
Parameters:
<PatternFrom>
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:PATLength <PatternLen>
Sets the word pattern length in bits. The maximal bit length of all selected pattern con-
ditions is 160 bits.
Suffix:
<m>
.
1..3
<n> 1..4
Selects the complex pattern condition.
Parameters:
<PatternLen> Range: 1 to 160
Increment: 1
*RST: 8
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:PCONdition <Operator>
Sets the operator to define the exact position or range where the instrument looks for
the specified pattern.
Suffix:
<m>
.
1..3
<n> 1..4
Selects the complex pattern condition.
Parameters:
<Operator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Trigger