Remote control commands
R&S
®
RTP
1287User Manual 1337.9952.02 ─ 12
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These operators require one identifier
pattern to be set with TRIGger<m>:SERPattern:COMPlex:
PATTern<n>:FRPTtern
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:SERPattern:COMPlex:
PATTern<n>:FRPTtern and TRIGger<m>:SERPattern:
COMPlex:PATTern<n>:TOPTtern.
*RST: EQUal
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:SRNGe
<SearchRange>
Sets the search range of the word pattern.
Suffix:
<m>
.
1..3
<n> 1..4
Selects the complex pattern condition.
Parameters:
<SearchRange> Range: 1 to 65535
Increment: 1
*RST: 1
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:TOPTtern <PatternTo>
Sets the the end value of a pattern range if TRIGger<m>:SERPattern:COMPlex:
PATTern<n>:PCONdition is set to INRange or OORange.
Suffix:
<m>
.
1..3
<n> 1..4
Selects the complex pattern condition.
Parameters:
<PatternTo>
22.9.13.4 PRBS error
TRIGger<m>:SERPattern:PRBS:ERATe.........................................................................1287
TRIGger<m>:SERPattern:PRBS:NBRTe.........................................................................1288
TRIGger<m>:SERPattern:PRBS:PSEQuence.................................................................1288
TRIGger<m>:SERPattern:PRBS:ERATe <PRBSBtErrRate>
Sets the bit error rate for the PRBS sequence.