EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 1393

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1393User Manual 1337.9952.02 ─ 12
MEASurement<m>:SPECtrum:CPOWer:BANDwidth <ChPowBw>
Defines the bandwidth over which the channel power is calculated.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ChPowBw> Range: 0 to 4E+9
Increment: 1
*RST: 0
Default unit: Hz
Usage: Asynchronous command
MEASurement<m>:SPECtrum:OBANdwidth <OccupiedBW>
Defines the percentage of the total power used to determine the occupied bandwidth.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<OccupiedBW> Range: 0.1 to 99.9
Increment: 1
*RST: 20
Default unit: %
Usage: Asynchronous command
MEASurement<m>:SPECtrum:CPOWer:CFRequency <ChPowCtrFreq>
Defines the center frequency from which the channel power is calculated over the
specified bandwidth.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ChPowCtrFreq> Range: 0 to 6E+9
Increment: 1
*RST: 0
Default unit: Hz
Usage: Asynchronous command
Automatic measurements

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals