TRIGger<m>:MILStd:DATA:DMIN <DataPatternMin>
Specifies a data pattern or sets the the start value of a data pattern range.
Parameters:
<DataPatternMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:DATA:ICONdition <DataIdxOperator>
Sets the operator to set a specific range within this series of the data words that is con-
sidered for the analysis.
Parameters:
<DataIdxOperator> EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
EQUal| LTHan | LETHan | GTHan | GETHan
Equal, Less than, Less or equal than, Greater Than, Greater or
equal than. These conditions require one pattern to be set with
TRIGger<m>:MILStd:DATA:IMIN.
INRange = RANGe
In range: set the minimum and maximum value of the range with
TRIGger<m>:MILStd:DATA:IMIN and TRIGger<m>:
MILStd:DATA:IMAX.
*RST: INRange
TRIGger<m>:MILStd:DATA:IMAX <DataIndexMax>
Sets the end value of a data word series index range if TRIGger<m>:MILStd:DATA:
ICONdition is set to INRange.
Parameters:
<DataIndexMax> Range: 1 to 32
Increment: 1
*RST: 32
TRIGger<m>:MILStd:DATA:IMIN <DataIndexMin>
Specifies an index or sets the the start value of a data word series index range.
Parameters:
<DataIndexMin> Range: 1 to 32
Increment: 1
*RST: 1