Remote control commands
R&S
®
RTP
1734User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMax>
TRIGger<m>:ETHernet:FRAMe:SMIN <SrcAddrPattMin>
Specifies a source address pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMin>
TRIGger<m>:ETHernet:FRAMe:TCONdition <TypeOperator>
Sets the type operator to trigger on a specific length/type value pattern or a range.
Suffix:
<m>
.
1..3
Parameters:
<TypeOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater Than,
Greater than or equal. These conditions require a destination
address to be set with TRIGger<m>:ETHernet:FRAMe:TMIN.
INRange | OORange
In range, Out of range. Set the minimum and maximum value of
the range with TRIGger<m>:ETHernet:FRAMe:TMIN and
TRIGger<m>:ETHernet:FRAMe:TMAX.
*RST: EQUal
TRIGger<m>:ETHernet:FRAMe:TMAX <TypePatternMax>
Sets the end value of a length/type range, if TRIGger<m>:ETHernet:FRAMe:
TCONdition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<TypePatternMax>
TRIGger<m>:ETHernet:FRAMe:TMIN <TypePatternMin>
Specifies a length/type value, or sets the start value of a pattern range.