Remote control commands
R&S
®
RTP
1863User Manual 1337.9952.02 ─ 12
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:DPHY:DIDX:CONDition <SearchName>,<Format>
SEARch:TRIGger:DPHY:DIDX:CONDition? <SearchName>
Sets the condition for the data index. You can define an exact value or a value range.
Parameters:
<Format> EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with the corresponding SEARch:TRIGger:DPHY:DIDX:
MIN command.
INRange = RANGe
In range: Set the minimum and maximum value using SEARch:
TRIGger:DPHY:DIDX:MIN and SEARch:TRIGger:DPHY:
DIDX:MAX.
*RST:
INRange
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:DPHY:DIDX:MAX <SearchName>,<DXSymbol>
SEARch:TRIGger:DPHY:DIDX:MAX? <SearchName>
Sets the the end value of a data index range if SEARch:TRIGger:DPHY:DIDX:
CONDition is set to INRange or RANge.
Parameters:
<DXSymbol> Range: 0 to 65535
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:DPHY:DIDX:MIN <SearchName>,<SymbolType>
SEARch:TRIGger:DPHY:DIDX:MIN? <SearchName>
Specifies a data index minimum, or sets the the start value of a range.
Parameters:
<SymbolType> Range: 0 to 65535
Increment: 1
*RST: 0
Protocols