Remote control commands
R&S
®
RTP
1886User Manual 1337.9952.02 ─ 12
Suffix: 
<m>
.
1..3
Parameters:
<FSNOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:MPHY:
FSNumber:MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MPHY:FSNumber:MIN and
TRIGger<m>:MPHY:FSNumber:MAX.
*RST: 
EQUal
TRIGger<m>:MPHY:FSNumber:MIN 
<FSNMin>
Specifies the FSN bit pattern, or sets the start value of a pattern range.
Suffix: 
<m>
.
1..3
Parameters:
<FSNMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:FSNumber:MAX <FSNMax>
Sets the end value of the FSN range if TRIGger<m>:MPHY:FSNumber:CONDition
is set to INRange or OORange.
Suffix: 
<m>
.
1..3
Parameters:
<FSNMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:CRC:CONDition <CRCOperator>
Sets the condition for CRC. You can define an exact value or a value range.
Protocols