Remote control commands
R&S
®
RTP
1950User Manual 1337.9952.02 ─ 12
Setting parameters:
<Frame>
<Field>
<Enabler> ON | OFF
Example:
:TRIGger:CMSB:ADVanced:FIEN "Master", "CRC", 1
Enables the CRC field of the Master frame.
:TRIGger:CMSB:ADVanced:FRAMe1:FLD3:ENABle 1
Enables the 3rd field of the 1st frame that you have defined.
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:BIT <BitState>
TRIGger<m>:CMSB:ADVanced:BIT <Frame>, <Field>, <Bit>
Sets the bit state of a field to be triggered on that only consists of one bit.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Bit> ONE | ZERO | DC
ONE
1
ZERO
0
DC
"Do not care" (DC) = X
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMAX <DataMax>
TRIGger<m>:CMSB:ADVanced:DMAX <Frame>, <Field>, <Data>
Sets the end value of a data pattern range if TRIGger<m>:CMSB:ADVanced:
DOPerator is set to INRange or OORange.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Data>