Remote control commands
R&S
®
RTP
1951User Manual 1337.9952.02 ─ 12
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMIN <DataMin>
TRIGger<m>:CMSB:ADVanced:DMIN <Frame>, <Field>, <Data>
Specifies the data pattern, or sets the start value of a data pattern range.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Data>
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DOPerator
<DataOperator>
TRIGger<m>:CMSB:ADVanced:DOPerator <Frame>, <Field>, <Operator>
Sets the operator for the data pattern.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Operator> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
ANY = OFF
The data of the required pattern is not relevant.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
This condition requires one endpoint value to be set using
TRIGger<m>:CMSB:ADVanced:DMIN.
INRange | RANGe
This condition requires a range of endpoint values to be set
using TRIGger<m>:CMSB:ADVanced:DMAX and
TRIGger<m>:CMSB:ADVanced:DMIN.
Usage:
Setting only
TRIGger<m>:CMSB:ADVanced:ERRor<n>:ENABle <Enable>
TRIGger<m>:CMSB:ADVanced:ERENable <ErrorName>, <Enabler>
Enables triggering on errors. You can search for all error types in parallel.