Remote control commands
R&S
®
RTP
2227User Manual 1337.9952.02 ─ 12
Suffix: 
<m>
.
1..2
Parameters:
<Frequency1> Range:  1E-3  to  100E+6
Increment:  1
*RST:  1E+6
Default unit: Hz
WGENerator<m>:MODulation:FSK:FTWO <Frequency2>
Sets the frequency of the modulated waveform, if WGENerator<m>:MODulation:
TYPE is set to FSK.
Suffix: 
<m>
.
1..2
Parameters:
<Frequency2> Range:  1E-3  to  100E+6
Increment:  1
*RST:  1000
Default unit: Hz
WGENerator<m>:MODulation:FSK[:RATE] <Rate>
Sets the hop rate, the time before a switch from the carrier frequency set with
WGENerator<m>:MODulation:FSK:FONE and the modulation frequency set with
WGENerator<m>:MODulation:FSK:FTWO.
Suffix: 
<m>
.
1..2
Parameters:
<Rate> Range:  1E-3  to  1E+6
Increment:  1
*RST:  1000
Default unit: Hz
WGENerator<m>:MODulation:PWM:DCYCle <SquareDutyCycle>
Sets the duty cycle of a square waveform for pulse width modulation, if
WGENerator<m>:MODulation:TYPE is set to PWM.
Suffix: 
<m>
.
1..2
Parameters:
<SquareDutyCycle> Range:  10  to  90
Increment:  1
*RST:  50
Default unit: %
Waveform generator (option R&S RTP-B6)