Remote control commands
R&S
®
RTP
2228User Manual 1337.9952.02 ─ 12
WGENerator<m>:MODulation:PWM:DEPTh <Depth>
Sets the modulation depth for pulse width modulation.
Suffix: 
<m>
.
1..2
Parameters:
<Depth> Range:  0  to  100
Increment:  1
*RST:  50
Default unit: %
WGENerator<m>:MODulation:PWM:FREQuency 
<Frequency>
Sets the frequency of the modulation waveform for pulse width modulation, if
WGENerator<m>:MODulation:TYPE is set to PWM.
Suffix: 
<m>
.
1..2
Parameters:
<Frequency> Range:  1E-3  to  1E+6
Increment:  1
*RST:  1000
Default unit: Hz
WGENerator<m>:MODulation:PWM:SYMMetry <RampSymmetry>
Sets the symmetry, the percentage of time the ramp modulation waveform is rising, for
pulse width modulation, if WGENerator<m>:MODulation:TYPE is set to PWM.
Suffix: 
<m>
.
1..2
Parameters:
<RampSymmetry> Range:  0  to  100
Increment:  1
*RST:  50
Default unit: %
WGENerator<m>:MODulation:PWM[:FUNCtion] <SignalType>
Sets the type of the modulation signal for the pulse width modulation, if
WGENerator<m>:MODulation:TYPE is set to PWM.
Suffix: 
<m>
.
1..2
Parameters:
<SignalType> SINusoid | SQUare | RAMP
*RST:  SINusoid
Waveform generator (option R&S RTP-B6)