359User Manual 1337.9952.02 ─ 12
"Level selection source 1/2": For each measurement source, select the reference level
on which the time is measured.
"Advanced Delay Setup":
With the settings shown in the picture, the time between the second and the fifth rising
edge is measured.
"Level selec-
tion"
Selects the reference level on which the time is measured.
"Slope"
Sets the edge of each source, between which the delay is measured:
positive, negative, or either of them.
"Direction"
Selects the direction for counting slopes for each source: from the
beginning of the waveform, or from the end.
"Number"
Sets the number of the edge that is relevant for delay measurement.
Remote command:
MEASurement<m>:AMPTime:DELay<n>:LSELect on page 1375
MEASurement<m>:AMPTime:DELay<n>:SLOPe on page 1376
MEASurement<m>:AMPTime:DELay<n>:DIRection on page 1375
MEASurement<m>:AMPTime:DELay<n>:ECOunt on page 1375
Settings for delay and phase measurements (digital sources)
Delay measurement on digital channels is reduced to measure the time between two
subsequent rising or two subsequent falling edges.
The edge direction is set with Edges slope.