Measurements
R&S
®
RTP
362User Manual 1337.9952.02 ─ 12
Trigger 
source
Measurement 
source
Trigger 
level
Reference 
level
Slope
Trigger 
event
Delay to trigger
To configure the trigger conditions, use the trigger setup. To set up the slope, additional
settings appear in the measurements "Setup" dialog box.
"Edge slope"
Sets the edge direction to be used for delay measurement: positive,
negative, or either edge.
"Level selec-
tion"
Selects the reference level of the measurement source on which the
delay is measured: upper, middle, or lower level.
"Signal threshold"
See "Signal threshold" on page 358.
Remote command: 
MEASurement<m>:AMPTime:DTOTrigger<n>:SLOPe on page 1378
MEASurement<m>:AMPTime:DTOTrigger<n>:LSELect on page 1378
8.2.5.3  Arithmetic for amplitude/time and spectrum measurements
Access: [Meas] > "Meas Group" tab > "Amp/Time"/"Spectrum" category > "Arithmetic".
In this dialog, you can set up basic arithmetic operations between selected measure-
ment values and/or constant.
You can select only measurement groups and measurements that are already enabled.
Automatic measurements