454User Manual 1337.9952.02 ─ 12
Polarity, Range, Width, ±Delta
See trigger settings:
"Polarity" on page 219
While the width trigger can only analyze positive or negative polarity, searching for
a width is also possible for both polarities at the same time ("Either").
"Range" on page 219
"Width" on page 220
"±Delta" on page 220
Remote command:
SEARch:TRIGger:WIDTh:POLarity on page 1494
SEARch:TRIGger:WIDTh:RANGe on page 1494
SEARch:TRIGger:WIDTh:WIDTh on page 1495
SEARch:TRIGger:WIDTh:DELTa on page 1494
Level
Sets the voltage level for trigger search. The value is used for all search event types
that require one trigger level. The search trigger level is search-specific, you can define
different levels in different searches for the same event.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1484
11.2.2.4 Runt
The runt search settings are the same as the runt trigger settings. To apply the trigger
settings to search, tap Copy from trigger A'.
The runt search is not available, if the search source is a digital channel.
Search setup