455User Manual 1337.9952.02 ─ 12
Polarity, Range, Runt width, ±Delta
Time limit for the runt, see trigger settings:
"Polarity" on page 218
"Range" on page 221
"Runt width" on page 222
"±Delta" on page 222
Remote command:
SEARch:TRIGger:RUNT:POLarity on page 1488
SEARch:TRIGger:RUNT:RANGe on page 1489
SEARch:TRIGger:RUNT:WIDTh on page 1489
SEARch:TRIGger:RUNT:DELTa on page 1488
Upper level, Lower level
Set the upper and lower voltage thresholds. The amplitude of a runt crosses the first
threshold twice in succession without crossing the second one.
Remote command:
SEARch:TRIGger:LEVel:RUNT:UPPer on page 1490
SEARch:TRIGger:LEVel:RUNT:LOWer on page 1490
11.2.2.5 Window
The window search settings are the same as the window trigger settings. This search
type is not available if the search source is a digital channel. To apply the trigger set-
tings to search, tap Copy from trigger A'.
The window search is not available if the search source is a digital channel.
Search setup