Protocol analysis
R&S
®
RTP
688User Manual 1337.9952.02 ─ 12
Triggers on "Master Read" command sequences. This sequence sends the register
address in the command payload, and reads two data frames.
You can refine the trigger condition:
Set a MID condition to trigger only on command sequences that are sent to the
specified MIDs.
See: "Slave Addr. /MID " on page 689.
Set a register address condition to trigger only on command sequences that write
data to the specified register.
See: "Address" on page 689.
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 690.
Master Context Transfer Write ← Type
Available only for "Version"= "2.1".
Triggers on "Master Context Transfer Write" command sequences.
You can refine the trigger condition:
Set an MID condition to trigger only on command sequences that are sent to the
specified MIDs.
See: "Slave Addr. /MID " on page 689
Set a byte count condition to trigger only on command sequences that send the
specified number of data frames.
See: "Byte count" on page 690
Set a register address condition to trigger only on command sequences that write
data to the specified register.
See: "Address" on page 689
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 690 and "Index" on page 690
Master Context Transfer Read ← Type
Available only for "Version"= "2.1".
Triggers on "Master Context Transfer Read" command sequences.
You can refine the trigger condition:
Set an MID condition to trigger only on command sequences that are sent to the
specified MIDs.
See: "Slave Addr. /MID " on page 689
Set a byte count condition to trigger only on command sequences that send the
specified number of data frames.
See: "Byte count" on page 690
Set a register address condition to trigger only on command sequences that write
data to the specified register.
See: "Address" on page 689
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 690 and "Index" on page 690
Error conditions ← Type
Triggers if at least one of the selected errors occurs in a frame. The instrument triggers
on the first detected error.
RFFE (option R&S
RTP-K40)