Protocol analysis
R&S
®
RTP
689User Manual 1337.9952.02 ─ 12
"SSC error"
Triggers when no valid SSC sequence has been found after the idle
time. The instrument triggers at the invalid sequence.
"Length error"
Triggers on an incorrect length of the command sequence - when
additional or missing bits are detected and the sequence of bits is not
as expected.
"No response"
Triggers on any No Response Frame. All 9 bits of a No Response
Frame, including the parity bit, are zero.
"Bus park error"
Triggers on an incorrect bus park cycle - when the instrument expects
a low bit in bus park but detects high bit.
"Parity error"
Triggers on incorrect parity bits.
"Unknown sequence"
Triggers when the instrument cannot detect any supported command
sequence.
"Gap error"
Triggers when timing gap error occurs.
"Version error"
Triggers on any version errors.
Remote command:
TRIGger<m>:RFFE:ERRor:SSC on page 1825
TRIGger<m>:RFFE:ERRor:LENGth on page 1824
TRIGger<m>:RFFE:ERRor:BP on page 1824
TRIGger<m>:RFFE:ERRor:PARity on page 1824
TRIGger<m>:RFFE:ERRor:NOResponse on page 1824
TRIGger<m>:RFFE:ERRor:USEQuence on page 1825
TRIGger<m>:RFFE:ERRor:GAP on page 1825
TRIGger<m>:RFFE:ERRor:VERSion on page 1825
Slave Addr. /MID
Defines the address of the slave / master ID. The slave address setup consists of the
condition and one or two address patterns.
"Condition"
Sets the operator to trigger on a specific address pattern ("Equal" or
"Not equal") or an address range.
"Slave Address
(Min)"
Defines the slave address pattern for all operators that require one
pattern.
"Slave Adress
(Max)"
Defines the second address pattern that is required to specify a range
with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:SADD:CONDition on page 1822
TRIGger<m>:RFFE:SADD:MIN on page 1823
TRIGger<m>:RFFE:SADD:MAX on page 1823
Address
Defines the register address. The register address setup consists of the condition and
one or two address patterns.
"Condition"
Sets the operator to trigger on a specific address pattern ("Equal" or
"Not equal") or an address range.