Protocol analysis
R&S
®
RTP
687User Manual 1337.9952.02 ─ 12
You can refine the trigger condition:
●
Set a slave address condition to trigger only on command sequences that are sent
to the specified MIDs.
See: "Slave Addr. /MID " on page 689
●
Set a bit count condition to trigger only on command sequences that send the
specified number of data frames.
See: "Byte count" on page 690
●
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 690
Masked Write ← Type
Available only for "Version"= "2.1".
Triggers on "Masked Write" command sequences. This sequence sends the register
address in the command payload, and sends one mask and one data frame.
You can refine the trigger condition:
●
Set a slave address condition to trigger only on command sequences that are sent
to the specified MIDs.
See: "Slave Addr. /MID " on page 689.
●
Set a register address condition to trigger only on command sequences that write
data to the specified register.
See: "Address" on page 689.
●
Set a mask pattern condition to trigger on a mask expected in the message.
See: "Mask" on page 691.
●
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 690.
Master Ownership Handover ← Type
Available only for "Version"= "2.0"/"2.1".
Triggers on "Master Ownership Handover" command sequences.
Optionally, you can specify an MID pattern or an MID range to trigger only on com-
mand sequences that are sent to these slaves.
Master Write ← Type
Available only for "Version"= "2.0"/"2.1".
Triggers on "Master Write" command sequences. This sequence sends the register
address in the command payload, and sends two data frames.
You can refine the trigger condition:
●
Set a MID condition to trigger only on command sequences that are sent to the
specified MIDs.
See: "Slave Addr. /MID " on page 689.
●
Set a register address condition to trigger only on command sequences that write
data to the specified register.
See: "Address" on page 689.
●
Set a data pattern condition to trigger on data patterns expected in the message.
See: "Data" on page 690.
Master Read ← Type
Available only for "Version"= "2.0"/"2.1".
RFFE (option R&S
RTP-K40)