5 - 12
IPN 074-397-P1G
HAPSITE Smart Operating Manual
5.2.1 Quick Reference SOP — Running Survey Mode
1 Double-click the Smart IQ software icon.
2 Double-click the Run Method icon.
3 Double-click the Survey Folder.
4 Double-click the Survey method.
5 Wait for heaters to reach the set temperatures.
6 Click the RUN button in the pop-up window, sample background for one minute.
7 Hold the Probe over the sample for one minute or until observing a response.
8 Press RUN again to stop the sampling and save the data.
HINT: This is a tentative identification. For more positive identification, run a
GCMS method.