9 - 1
IPN 074-397-P1G
HAPSITE Smart Operating Manual
Chapter 9
Data Review
9.1 Introduction to Data Review
The Data Review section of the HAPSITE SMART software allows access to
previously acquired data for review and analysis, or to view data that is being
acquired in real time. Data Review functionality includes:
Search quickly for individual components using the NIST Library and the F7
key.
Search for Peaks function to scan and tentatively identify all peaks in a Total
Ion Chromatogram (TIC), as well as generate a report.
Specify which masses to view in a Reconstructed Ion Chromatogram (RIC)
Select Scan ranges to be viewed in the TIC.
Zoom in/out of TIC/RIC.
Subtract backgrounds from TIC.
Subtract specific scans from acquired spectra.
Grab spectra for analysis utilizing the full NIST database program.
Go to AMDIS (Automated Mass Spectral Deconvolution and Identification
System) with TIC for qualitative identification.
Review data file information.
Label TIC and individual spectra.
View the method used to acquire data.
View the GC column temperature profile (if applicable).
Adjust properties of the Data Review (Monitor).
These features will allow the tentative identification of unknowns using the NIST or
AMDIS spectral libraries, and prepare reports that will aid in the identification of
unknown chemicals.