1 - 2
IPN 074-397-P1G
HAPSITE Smart Operating Manual
1.2 Performance Specifications
The performance specifications for the HAPSITE are shown below:
Mass range . . . . . . . . . . . . . . . . . . 1-300 AMU
Scan Rate . . . . . . . . . . . . . . . . . . . as much as 1000 AMU/sec @ 10 points per
AMU
Ionization Mode . . . . . . . . . . . . . . . 70 eV electron impact
Vacuum System . . . . . . . . . . . . . . . 15 l/sec NEG pump;
0.2 l/sec sputter-ion pump
Operating temperature range. . . . . 0
o
C to 45
o
C (32 to 113
o
F)
Dimensions (LxWxH) . . . . . . . . . . . 46 x 43 x 18 (cm); 18 x 17x 7 (in)
Weight . . . . . . . . . . . . . . . . . . . . . . 16 Kg (35 lb.) without the battery
Internal Power Consumption . . . . . 30 watts average, 24 V(dc)
Carrier Gas . . . . . . . . . . . . . . . . . . Nitrogen
Column Temperature Range . . . . . 60
o
C to 180
o
C
Maximum Sample Moisture Content 8% by weight
pH Range of Sample . . . . . . . . . . . . 2 - 11
Boiling Point of Sample . . . . . . . . . . <250
o
C
Chemical Composition of Sample . . 1 - 12 Carbon atoms
GC Column . . . . . . . . . . . . . . . . . . 100% methyl silicone phase,
30 m x 0.32 mm ID x 1.0 film
SIM Channels . . . . . . . . . . . . . . . . 10
External Communications . . . . . . . Ethernet Port
Carrier gas use-rate. . . . . . . . . . . .1 canister per 8 hours of operation
(depending on the details of the method
being used).
Internal standards gas use-rate . . . 1 canister per 24 hours of operation
(depending on the details of the method
being used).
Battery life . . . . . . . . . . . . . . . . . . . Approximately 2 to 3 hours before
recharge.
1.3 Serial Number Location
The serial number of the HAPSITE is located on the inside of the front panel under
the power switch.