15.6.2 Reactivating the NEG Pump
Reactivating the NEG Pump requires having the Service Module attached to the
HAPSITE for at least 22 hours. The procedure is basically the same as activating
a new NEG pump. Refer to section 2.7, Service Module, on page 2-18 for more
information on setting up the Service Module.
1 Click on the Service Module tab of the Properties menu.
2 Click on the Service button. See Figure 15-10.
Figure 15-10 Service Button on the Service Module Tab
3 Use the Activate NEG settings below as a guideline except change HAPSITE
Pumpdown Time to 180. See Figure 15-11.