3 - 6
IPN 074-397-P1G
HAPSITE Smart Operating Manual
3.1.1 Quick Reference SOP - Heat-up and Tune
1 Insert the Internal Standard and Carrier gas canisters.
2 Insert a charged battery
3 Connect the AC to DC power converter power supply.
4 Press the power button on the inside of the front panel.
NOTE: If a startup method is active, the HAPSITE will automatically perform
Step 5 on page 3-6 through Step 7 on page 3-6.
HINT: If the wireless connection to the laptop computer is to be used, see
Chapter 4, Wireless and USB.
5 From the Main Menu highlight Run Method and press SEL.
6 Highlight either Survey or GCMS and press SEL (depends on what is being
run). If choosing the GCMS, highlight and select the folder desired.
7 Highlight the desired method (Air 15 min Loop or Survey)
NOTE: The HAPSITE will now heat up the all zones and perform a tune check.
A run method prompt will appear when the HAPSITE is ready to run a
sample. The process for heat-up and tune takes approximately 20
minutes.
3.2 HAPSITE Front Panel Main Menu
The folder tree on the HAPSITE is a menu of operations that can be performed by
pressing buttons on the front panel of the HAPSITE.
In all cases, pressing the SEL button will select the currently highlighted selection.
Pressing the ESC button moves back to the previous menu.
Figure 3-7 shows the HAPSITE front panel and LCD Main Menu display.
NOTE: The blue banner at the top of the Front Panel does not exist on the actual
Front Panel (this is a computer representation of the front panel which is
displayed on the HAPSITE computer software). Also, the LCD button is
only on the computer representation.