5 - 31
IPN 074-397-P1G
HAPSITE Smart Operating Manual
5.5.1 Quick Reference SOP — Tenax Cleanout
1 Double-click the Smart IQ software icon.
2 Double-click on the Run Method icon.
3 Double-click the GCMS folder.
4 Double-click the Concentrator folder.
5 Double-click the 3 min Tenax Conc Cleanout method.
6 Click the RUN button displayed in the pop-up window on the laptop. This run
will take three minutes to complete.
NOTE: The TIC_Max number must be less than 500,000 for the concentrator tube
to be considered clean. If the TIC_Max number is greater than 500,000,
repeat the 3 min Tenax Conc Cleanout.mth method until the TIC_Max
number is less than 500,000.
5.5.2 Quick Reference SOP — Tenax Concentrator Method
NOTE: The 3 min Tenax Conc Cleanout method must be run prior this method.
Refer to Section 5.5.1, Quick Reference SOP — Tenax Cleanout, on page
5-31.
1 Double-click the Smart IQ software icon.
2 Double-click on the Run Method icon.
3 Double-click the GCMS folder.
4 Double-click the Concentrator folder.
5 Double-click the Air 15 min Tenax Conc method.
6 Wait for heaters to gain control of set temperatures.
7 Click Run on the pop-up window in the software.
8 Hold the Probe over the sample for the entire Linepurge and Concfill events.
This takes two minutes.
9 Review results when the run is completed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon HAPSITE Smart and is the answer not in the manual?

Inficon HAPSITE Smart Specifications

General IconGeneral
BrandInficon
ModelHAPSITE Smart
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals