5 - 23
IPN 074-397-P1G
HAPSITE Smart Operating Manual
11 Highlight the Air 15 min Carbon Conc and then press OK. See Figure 5-37.
Figure 5-37 Selecting the Air 15 min Carbon Conc method
12 The software will check the pressures and temperatures and prompt when
ready to run a sample. See Figure 5-38.
Figure 5-38 Air 15 min Carbon Conc Run Prompt
13 Hold the Probe over the sample and press RUN.
CAUTION
Do not place the sample probe in liquids while sampling.
14 The sampling process takes two minutes — one minute to flush the system and
one minute to collect the sample. The probe needs to be held over the sample
for the entire two minutes while the Sample Pump is running. The bottom left
corner will say LinePurge, then Concfill on the Front Panel display.