HAPSITE Smart Operating Manual
Running Tune Base Peak Gain . . . . . . . . . .Current measured Base Peak Gain.
NOTE: The base peak gain will switch to
Red when BPG is outside the target
range.
Auto Resolve . . . . . . . . . . . . . . . . . Adjusts the resolution of all Mass Peaks to
the target resolution.
Save Tune . . . . . . . . . . . . . . . . . . . . Save the Tune File
Mass Cal.. . . . . . . . . . . . . . . . . . . . . Checks and corrects the ten calibration
masses for correct location within the mass
range.
Full Scan . . . . . . . . . . . . . . . . . . . . . Switches between Full Scan display mode
and Peak Scan display mode
Short Auto Tune . . . . . . . . . . . . . . . Starts the Short Autotune function
Long Auto Tune . . . . . . . . . . . . . . . Starts the Long Autotune function
Noise Check . . . . . . . . . . . . . . . . . . Scans a "quiet" mass range (no peaks) to
determine the electronic baseline and
threshold noise level.
Tune Checkup. . . . . . . . . . . . . . . . . Runs a mass calibration and noise check.
Zoom . . . . . . . . . . . . . . . . . . . . . . . . Enables the cursor to select a section of Full
scan or a peak scan and zoom to that
section.
Mass Adjust . . . . . . . . . . . . . . . . . . Enables the cursor to select a mass peak
and drag the mass peak to a different
position on the mass axis.