13 - 7
IPN 074-397-P1G
HAPSITE Smart Operating Manual
1 Rinse a small clean vessel (such as a beaker) with VOC-free water and discard
the water.
2 Pour some VOC-free water in the rinsed vessel for rinsing of the sample
syringe.
3 Rinse the sample syringe with VOC-free water and discard the water. Repeat
three times.
4 Remove the plunger from the sample syringe. Hold the sample syringe at an
angle with the large open end up, and the small end (where a needle would be)
stopped against a gloved finger. See Figure 13-4. Carefully transfer the sample
into the sample syringe.
Figure 13-4 Transferring the Sample into a Sample Syringe
NOTE: The 40 mL sample vials are intended for one time use only and should
be discarded properly when finished.
5 Adjust the amount of sample in the sample syringe to 20 mL while removing the
air bubble. See Figure 13-5.
WARNING
Do not insert a vial which is completely full of liquid
sample. An adequate airspace above the sample is
required to allow for sample expansion during heating.
Failure to leave an airspace in the vial may result in
failure of the vial and/or personal injury. Do not fill the vial
with more then 20mL of liquid.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon HAPSITE Smart and is the answer not in the manual?

Inficon HAPSITE Smart Specifications

General IconGeneral
BrandInficon
ModelHAPSITE Smart
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals