15 - 42
IPN 074-397-P1G
HAPSITE Smart Operating Manual
The NEG Pump becomes very hot when in operation. The
software-controlled sequence to remove the pump
provides a cool down period time to cool. This
sequencing must not be bypassed. In addition to the
danger of burns from handling the hot flange, the
reactive metal alloys within the pump can ignite if
exposed to atmospheric pressure while hot. If this should
occur, drop the pump into a bucket of sand and cover the
hot NEG Pump with more sand. If this is not practical, set
the pump on the black plastic base, on a non-flammable
surface. The NEG Pump will burn out in a few minutes.
4 When the screen shows Vent Manifold Has Been Completed, unplug the
NEG Pump’s black cable from the front panel by loosening the locking ring,
then pull the connector straight back.
Do not disconnect the NEG Pump before the Vent
Manifold Has Been Completed message, as the
temperature sensor would then be inaccessible to the
controller.