10 SENSe Subsystem
356 Keysight N8262A Programming Guide
Query Example
Error Messages
If [SENSe[1]]|SENSe2:AVERage:STATe is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings Conflict”
occurs.
SENS2:AVER? This command queries whether averaging is
on or off for Channel B.