SENSe Subsystem 10
Keysight N8262A Programming Guide 369
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
This command enters the name of the sensor calibration table or frequency
dependent offset table which is to be used. The CSET1 command selects the
sensor calibration table and the CSET2 command selects the frequency dependent
offset table. The calibration factor is interpolated from the table using the setting
for [SENSe[1]]|SENSe2:FREQuency.
Syntax
Parameters
NOTE
If [SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe is set to OFF, the
selected sensor calibration table or frequency offset table is not being used.
Item Description/Default Range of Values
string String data representing a sensor calibration table, or
frequency dependent offset table name.
Any existing table name
(Existing table names
can be listed using MEMory:CATalog:TABle?).
SENS
1
:CSET
CORR
SENS2
:
?
:SEL
string
Space
1
:CSET2

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals