3 WLAN Mode
3.7 WLAN MIMO Modulation Analysis Measurement
3.7.9.15 MIMO EVM Filter Short State (Remote Command Only)
This setting takes effect when MIMO EVM Multi-Burst is On. It is used to filter out the
spurs in the Multi-Burst acquisition data.
Remote Command
[:SENSe]:EVMMimo:EVM:MBURst:FILTer 0 | 1 | Off | On
[:SENSe]:EVMMimo:EVM:MBURst:FILTer?
Notes When MIMO EVM Multi-Burst State and this setting are both On, the width of the first burst captured
in the acquisition is used as standard burst width. If a following burst width is less than that of the 1st
burst. The shorter burst data will not be demodulated
Dependencies This setting takes effect when MIMO EVM Multi-Burst is On
Preset 0
3.7.10 Sweep
The Sweep key accesses controls that enable you configure and control the
acquisition of data and the X-axis parameters of the instrument. These controls
might include Continuous/Single, Pause/Resume, X Scale and number of Points.
3.7.10.1 Sweep/Control
Accesses controls that let you operate the sweep and control functions of the
instrument, such as Sweep Time and Continuous/Single.
Sweep/Measure
Lets you toggle between Continuous and Single sweep or measurement operation.
The single/continuous state is Meas Global, so the setting affects all measurements.
The front-panel key Single/Cont performs exactly the same function
See "More Information" on page 1296
Remote
Command
:INITiate:CONTinuous OFF | ON | 0 | 1
:INITiate:CONTinuous?
Example Put instrument into Single measurement operation:
:INIT:CONT 0
:INIT:CONT OFF
Put instrument into Continuous measurement operation:
:INIT:CONT 1
:INIT:CONT ON
1295 WLAN Mode User's &Programmer's Reference