Keysight M9421A - Restore Defaults; 3.3.9 Sweep; 3.3.9.1 Sweep;Control; Sweep Time

Keysight M9421A
2667 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 WLAN Mode
3.3 Occupied BW Measurement
Remote Command
:INSTrument:COUPle:EMC:STANdard ALL | NONE
:INSTrument:COUPle:EMC:STANdard?
Example
:INST:COUP:EMC:STAN ALL
:INST:COUP:EMC:STAN?
Dependencies Only available if Option EMC is installed
Preset Set to OFF on Global Settings, Restore Defaults and System, Restore Defaults, All Modes
Range
ALL | NONE
Restore Defaults
Resets all functions in the Global settings menu to OFF. Pressing System, Restore
Defaults, All Modes has the same effect.
Remote Command
:INSTrument:COUPle:DEFault
Example
:INST:COUP:DEF
Backwards Compatibility
SCPI
:GLOBal:DEFault
3.3.9 Sweep
Accesses controls to configure and control the acquisition of data, and the X-axis
parameters of the instrument.
Depending on the selected mode and measurement, these controls might include:
Sweep Time, Continuous/Single, Pause/Resume, X Scale and Number of Points.
3.3.9.1 Sweep/Control
Accesses controls that let you operate the sweep and control functions of the
instrument, such as Sweep Time and Continuous/Single.
Sweep Time
Controls the time the instrument takes to sweep the current frequency span in
swept measurements, displays the sweep time in swept measurements, and
displays the equivalent Sweep Time in FFT measurements.
When Sweep Time is in Auto, the instrument computes a time that will give accurate
measurements based on other settings, such as RBW and VBW.
423 WLAN Mode User's &Programmer's Reference