3 WLAN Mode
3.7 WLAN MIMO Modulation Analysis Measurement
2223 under Input/Output.
This menu includes the following display-only fields:
Sample Rate
Displays the sample rate of the saved IQ data file. If you have not saved an IQ data
file, the value is 0.
Sample Points
Displays the total number of sample points in the saved IQ data file. If you have not
saved an IQ data file, the value is 0.
Sampling Time
Displays the total sample time of the saved IQ data file. If you have not saved an IQ
data file, the value is 0.
Saved Channels [Mode: 5GNR, VMA, WLAN]
Displays the channel index/number of the saved IQ data file. If you have not saved
an IQ data file, no value is displayed.
This control is designed for multi-channel I/Q data recording and playback.
3.7.10.4 Playback
Contains parameters for playback of saved recording files recalled to the
instrument.
Recording and playback of signal data files is a multi-step process that involves
controls in several menus:
Save, Recording (under the Save hardkey or the Save icon in the File panel)
Recall, Recording (under the Recall hardkey or the Recall icon in the File panel)
Sweep, "Recording" on page 1535
WLAN Mode User's &Programmer's Reference 1304