3 WLAN Mode
3.8 Power vs Time Measurement
Fall: 3
Example
:DISP:PVT:WIND:TRAC:Y:SCAL:RLEV 5
:DISP:PVT:WIND:TRAC:Y:SCAL:RLEV?
Couplings When Auto Scaling is On (default), this value is automatically determined by the measurement result.
When you set a value manually, Auto Scaling changes to Off. Attenuation is not coupled to Ref Value.
Preset 10.00 dBm ! 1-Burst
0.00 dBm ! 2-Rise
0.00 dBm ! 3-Fall
State Saved Saved in instrument state
Backwards Compatibility SCPI
Window Numbers used to be a combination of View and Window, now the VIEW
node is not used and defaults to 1. All Windows are referenced through VIEW1. For
backwards compatibility the old View and Window numbers are honored as below:
Window Old SCPI Command New SCPI Command
PVT -
Burst
:DISP:PVT:VIEW
[1]:WIND:TRAC:Y:SCAL:RLEV 5dBm
:DISP:PVT:WIND1:TRAC:Y:SCAL:RLEV
5dBm
PVT – Rise :DISP:PVT:VIEW2:WIND:TRAC:Y:SCAL:RLEV
5dBm
:DISP:PVT:WIND2:TRAC:Y:SCAL:RLEV
5dBm
PVT – Fall :DISP:PVT:VIEW2:WIND2:TRAC:Y:SCAL:RLE
V 5dBm
:DISP:PVT:WIND3:TRAC:Y:SCAL:RLEV
5dBm
Scale/Div
Scale/Div sets the height of one division of the graticule in the current Y-Axis unit.
Scale/Div also determines the displayed amplitude range in the log plot graph.
Since there are usually 10 vertical graticule division on the display, the total
amplitude range of the graph is typically 10x this amount. For example, if Scale/Div
is 10 dB, then the total range of the graph is 100 dB.
Remote Command
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:Y[:SCALe]:PDIVision <rel_ampl>
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:Y[:SCALe]:PDIVision?
Window numbers are as follows:
Burst: 1
Rise: 2
Fall: 3
Example
:DISP:PVT:WIND:TRAC:Y:PDIV 5
1-Burst for Transmit On/Off Power
WLAN Mode User's &Programmer's Reference 1314