3 WLAN Mode
3.9 Spectral Flatness Measurement
All Markers Off
Turns off all markers.
Remote
Command
:CALCulate:FLATness:MARKer:AOFF
Example
:CALC:FLATness:MARK:AOFF
Couple Markers
When this function is On, moving any marker causes an equal X Axis movement of
every other marker which is not Off. “equal X Axis movement” means to preserve the
difference between each marker’s X Axis value (in the fundamental x-axis units of
the trace that marker is on) and the X Axis value of the marker being moved (in the
same fundamental x-axis units).
Remote
Command
:CALCulate:FLATness:MARKer:COUPle[:STATe] ON | OFF | 1 | 0
:CALCulate:FLATness:MARKer:COUPle[:STATe]?
Example
:CALC:FLATness:MARK:COUP ON
:CALC:FLATness:MARK:COUP?
Notes In general, when coupling is turned on then all Normal or Delta markers with the same (or equivalent)