9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Bit Condition Operation
signal or settings conditions
8 Signal Too
Noisy
The instrument cannot measure the desired signal because it is too
noisy
9 Slot Error No valid signal slot found in captured data
Filter Registers
"Questionable Integrity Signal Condition" on page 2639
"Questionable Integrity Signal Enable" on page 2639
"Questionable Integrity Signal Event Query" on page 2640
"Questionable Integrity Signal Negative Transition" on page 2640
"Questionable Integrity Signal Positive Transition" on page 2641
Questionable Integrity Signal Condition
Returns the decimal value of the sum of the bits in the Questionable Integrity Signal
Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote
Command
:STATus:QUEStionable:INTegrity:SIGNal:CONDition?
Example
:STAT:QUES:INT:SIGN:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Questionable Integrity Signal Enable
Determines which bits in the "Questionable Integrity Signal Condition" on page 2639
Register will set bits in the "Questionable Integrity Signal Event Query" on page
2640 register, which also sets the Integrity Summary bit (bit 9) in the "Questionable
Register" on page 2609.
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
:STATus:QUEStionable:INTegrity:SIGNal:ENABle <integer>
2639 WLAN Mode User's &Programmer's Reference