EasyManua.ls Logo

Keysight M9421A - Marker Settings Diagram; All Markers Off; Couple Markers; 3.12.7.3 Peak Search

Keysight M9421A
2667 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 WLAN Mode
3.12 IQ Waveform Measurement
marker is already a Delta marker, the reference marker is moved to the current
position of the selected marker, thus resetting the delta to zero.
Marker Settings Diagram
Lets you configure the Marker system using a visual utility.
All Markers Off
Turns off all markers.
Remote
Command
:CALCulate:WAVeform:MARKer:AOFF
Example
:CALC:WAV:MARK:AOFF
Couple Markers
When this function is ON, moving any marker causes an equal X-Axis movement of
every other marker that is not OFF. By “equal X-Axis movement” we mean that we
preserve the difference between each marker’s X-Axis value (in the fundamental x-
axis units of the trace that marker is on), and the X-Axis value of the marker being
moved (in the same fundamental x-axis units).
This may result in markers going off-screen.
Remote
Command
:CALCulate:WAVeform:MARKer:COUPle[:STATe] ON | OFF | 1 | 0
:CALCulate:WAVeform:MARKer:COUPle[:STATe]?
Example
:CALC:WAV:MARK:COUP ON
:CALC:WAV:MARK:COUP?
Preset
OFF
Presets on Mode Preset and "All Markers Off" on page 1817
State Saved Saved in instrument state
3.12.7.3 Peak Search
The controls on this tab allow you to move the marker to selected peaks of the
signal, giving you enormous analysis capabilities, particularly when combined with
the Delta marker function.
NOTE
Pressing the Peak Search hardkey automatically moves you to the Peak Search
page of the Marker menu
and
performs a peak search.
1817 WLAN Mode User's &Programmer's Reference