9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
"Questionable Calibration Extended Needed Negative Transition" on page 2627
"Questionable Calibration Extended Needed Positive Transition" on page 2627
Questionable Calibration Extended Needed Condition
Returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Needed Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?
Example
:STAT:QUES:CAL:EXT:NEED:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Extended Needed Enable
Determines which bits in the "Questionable Calibration Extended Needed Condition"
on page 2626 will set bits in the "Questionable Calibration Extended Needed Event
Query" on page 2627 register, which also sets bit 14 of the "Questionable
Calibration Register" on page 2621.
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle <integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?
Example
:STAT:QUES:CAL:EXT:NEED:ENAB 2
Can be used to query if an EMI conducted alignment is needed
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
WLAN Mode User's &Programmer's Reference 2626