3 WLAN Mode
3.11 Monitor Spectrum Measurement
[:SENSe]:MONitor:SWEep:POINts?
Example
:MON:SWE:POIN 1000
:MON:SWE:POIN?
Dependencies This function is not available when Signal ID is set to On in External Mixing
Neither the knob nor the step keys can be used to change this value. If it is tried, a warning is given
Not displayed in modes that do not support Swept
Couplings Whenever the number of sweep points change:
–
All trace data is erased
–
Any traces with Update Off will also go to Display Off (like going from View to Blank in the older
instruments)
–
Sweep time is re-quantized
–
Any limit lines that are on will be updated
–
If averaging/hold is on, averaging/hold starts over
–
Auto Sweep Points is OFF (Only 5GNR)
The resolution of setting the sweep time depends on the number of points selected
Preset 1001 unless noted below
2001: 5GNR
State Saved Saved in instrument state
Min 1
Max 20001
Annotation On second line of annotations, in lower right corner in parenthesis behind the sweep annotation
3.11.10 Trace
Lets you control the acquisition, display, storage, detection, and manipulation of
trace data for the available traces.
The "Trace Control" on page 1728 tab in this menu contains radio-button selections
for the trace type (Clear/Write, Trace Average, Max Hold, Min Hold) and View/Blank
setting for the selected trace.
3.11.10.1 Select Trace
Specifies the selected trace, which is the trace that will be affected when you
change trace settings.
1727 WLAN Mode User's &Programmer's Reference