9 Programming the Instrument
9.2 IEEE 488.2 Common Commands
9.2 IEEE 488.2 Common Commands
The instrument supports the following subset of IEEE 488.2 Common Commands, as
defined in Chapter 10 of IEEE Standard 488.2–1992. As indicated below, some of
these commands correspond directly to instrument front-panel functionality, while
others are available only as remote commands.
"*CAL? - Calibration Query" on page 2559 (Align Now All equivalent)
"*CLS - Clear Status" on page 2560
"*ESE - Standard Event Status Enable" on page 2560
"*ESR? - Standard Event Status Register Query" on page 2561
"*IDN? - Identification Query" on page 2561
"*OPC? - Operation Complete" on page 2562
"*OPT? - Query Instrument Options" on page 2563
"*RCL - Recall Instrument State" on page 2563 (Recall State equivalent)
"*RST - Reset" on page 2563 (Mode Preset equivalent)
"*SAV - Save Instrument State" on page 2564 (Save State equivalent)
"*SRE - Service Request Enable" on page 2564
"*STB? - Status Byte Query" on page 2565
"*TRG - Trigger" on page 2565
"*TST? - Self Test Query" on page 2565
"*WAI - Wait-to-Continue" on page 2566
9.2.1 *CAL? - Calibration Query
*CAL? Performs a full alignment and returns a number indicating the success of the
alignment. A zero is returned if the alignment is successful. A one is returned if any
part of the alignment fails. The equivalent SCPI command is :CALibrate[:ALL]?
See "Align Now All" on page 1927
Remote Command
*CAL
2559 WLAN Mode User's &Programmer's Reference