3 WLAN Mode
3.12 IQ Waveform Measurement
Example
:WAV:AVER:TAC 10
:WAV:AVER:TAC?
Notes Only available when Option DP2, B40, or wider IF Bandwidth option is installed
Preset 1
State Saved Saved in instrument state
Min/Max 1/65535
Auto Function
Remote
Command
[:SENSe]:WAVeform:AVERage:TACount:AUTO OFF | ON | 0 | 1
[:SENSe]:WAVeform:AVERage:TACount:AUTO?
Example
:WAV:AVER:TAC:AUTO ON
:WAV:AVER:TAC:AUTO?
Preset
ON
Range
Auto|Man
Meas Time
Lets you set how long the measurement is performed. X Scale only changes the
scale of the display.
Remote
Command
[:SENSe]:WAVeform:SWEep:TIME <time>
[:SENSe]:WAVeform:SWEep:TIME?
Example
:WAV:SWE:TIME 50 ms
:WAV:SWE:TIME?
Notes Specifies and returns how long the measurement is performed. It is the time record length of the
measurement waveform. The Max time may be reduced when the sample frequency is high due to the
memory limitation
Preset LTE, LTETDD, LTEAFDD, LTEATDD, 5G NR Modes: 10 ms
All other Modes: 2.000000 ms
State Saved Saved in instrument state
Range 1.000 us to 100.00 s
Min/Max 1.000 us/100.0 s
1.000 us/3200 s
Sample Rate
Sets an arbitrary sample rate for the acquired data to be processed.
Remote Command
[:SENSe]:WAVeform:SRATe <freq>