EasyManua.ls Logo

Keysight M9421A - Pause;Resume; 3.8.8.2 X Scale; Ref Value

Keysight M9421A
2667 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 WLAN Mode
3.8 Power vs Time Measurement
The displayed average/hold, K, shows the count for the sweep (data acquisition) in
progress. K = k + 1, with a limit of N. The displayed value K changes from its previous
value to 1 as soon as the trigger condition for the first data acquisition (sweep) is
met.
Pause/Resume
Pauses a measurement after the current data acquisition is complete.
When paused, the label on the control changes to Resume. Pressing Resume un-
pauses the measurement. When paused, pressing Restart performs a Resume.
Remote Command
:INITiate:PAUSe
:INITiate:RESume
Example
:INIT:PAUS
:INIT:RES
Dependencies Not displayed in Modes that do not support pausing
Annotation Only on control
3.8.8.2 X Scale
This tab accesses controls that enable you to configure the Sweep and Control
functions of the analyzer, such as Sweep Rules.
Ref Value
Enables you to set the display X reference value.
Remote
Command
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:X[:SCALe]:RLEVel <time>
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:X[:SCALe]:RLEVel?
Window numbers are as follows:
Burst: 1
Rise: 2
Fall: 3
Example
:DISP:PVT:WIND:TRACE:X:RLEV 1s
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND:TRACE:X:RLEV?
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND2:TRAC:X:RLEV 1s
2-Rise for Transmit On/Off Power
:DISP:PVT:WIND3:TRAC:X:RLEV 1s