3 WLAN Mode
3.8 Power vs Time Measurement
Burst: 1
Rise: 2
Fall: 3
Example
:DISP:PVT:WIND:TRAC:Y:COUP ON
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND:TRAC:Y:COUP? !
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND2:TRAC:Y:COUP ON
2-Rise for Transmit On/Off Power
:DISP:PVT:WIND3:TRAC:Y:COUP?
2-Fall for Transmit On/Off Power
Couplings When Auto Scaling is On, and the Restart front-panel key is pressed, this function automatically sets
the scale per division to 10 dB and determines the reference values based on the measurement results.
When you set a value to either Scale/Div or Ref Value manually, Auto Scaling automatically changes to
Off.
Preset 1
State Saved Saved in instrument state.
Range On|Off
Backwards Compatibility SCPI
Window Numbers used to be a combination of View and Window, now the VIEW
parameter is not used and defaults to 1. All Windows are referenced through
VIEW1. For backwards compatibility the old View and Window numbers are
honored as below:
Window Old SCPI Command New SCPI Command
PVT - Burst :DISP:PVT:VIEW[1]:WIND:TRAC:Y:COUP ON :DISP:PVT:WIND[1]:TRAC:Y:COUP
ON
PVT – Rise :DISP:PVT:VIEW2:WIND:TRAC:Y:COUP ON DISP:PVT:WIND2:TRAC:Y:COUP ON
PVT – Fall :DISP:PVT:VIEW2:WIND2:TRAC:Y:COUP ON DISP:PVT:WIND3:TRAC:Y:COUP ON
3.8.2.2 Attenuation
Controls the attenuator functions and interactions between the attenuation system
components.
There are two attenuator configurations in the X-Series. One is a Dual-Attenuator
configuration consisting of a mechanical attenuator and an optional electronic
attenuator. The other configuration uses a single attenuator with combined
1317 WLAN Mode User's &Programmer's Reference